EPMD, Eclipse Prism Medical Devices
Triple Dome System
Luminox Series

Triple Dome System

Triple-head ceiling system with one Luminox 300 and two Luminox 500 heads for redundancy and teaching theatres.

Triple-dome Luminox systems deliver redundant illumination and broad coverage for complex rooms and teaching theatres.

This configuration combines one Luminox 300 head with two Luminox 500 heads for compact auxiliary fields and variable focus on the primary positions.

Key highlights

  • Triple dome configuration for broad, redundant coverage
  • One Luminox 300 and two Luminox 500 heads
  • 100,000 lux (300) and 160,000 lux (500) per head at 1 m
  • Variable Narrow and Broad focus on Luminox 500 heads
  • 60,000+ hour LED service life

Specifications

Product specifications
ParameterDome 1: Luminox 300Dome 2: Luminox 500Dome 3: Luminox 500
Max. Central illumination at 1 m100,000 lux160,000 lux160,000 lux
Light field diameter at 1 m (d10)250 mm ± 10%200 mm - 300 mm ± 10%200 mm - 300 mm ± 10%
Light head diameter300 mmup to 600 mmup to 600 mm
Depth of illumination at 20% (L1+L2)1300 mm1200 mm1200 mm
Depth of illumination at 60% (L1+L2)800 mm750 mm750 mm
Correlated Color Temperature (CCT)6500 K ± 500 K3500 K - 5500 K3500 K - 5500 K
Color Rendering Index (CRI)> 90> 92> 92
Power consumption< 25 W< 50 W< 50 W
Focus ModeFixedVariable (Narrow, Broad)Variable (Narrow, Broad)
Dimming Range0 – 100 %0 – 100 %0 – 100 %
No. of LEDs123030
LED Service Life60,000 hours60,000 hours60,000 hours
Supply Voltage230 V ±10%, 50 Hz ±5%230 V ±10%, 50 Hz ±5%230 V ±10%, 50 Hz ±5%
Power Supply Unit24 V DC ±10%, 4.2 A max24 V DC ±10%, 4.2 A max24 V DC ±10%, 4.2 A max
Operating Temperature5°C - 40°C5°C - 40°C5°C - 40°C
Humidity10% - 90%10% - 90%10% - 90%
  • The specifications mentioned above can be used for wall, single, double, triple and mobile versions of Luminox series.
  • All values conform to IEC 60601-2-41 and are subject to a general industrial tolerance ± 5%.
  • In accordance with IEC 60601-2-41, the maximum illuminance must not exceed 160,000 lux.
  • Focus Mode: Variable = Narrow and Broad mode (Luminox 500). Luminox 300 uses fixed focus.
  • Custom development available for programme-specific mount and interface requirements.