EPMD, Eclipse Prism Medical Devices
Double Dome System
Luminox Series

Double Dome System

Dual-head ceiling system with Luminox 300 and Luminox 500 for shadow control across wider fields.

Dual-dome Luminox configurations provide overlapping fields and improved shadow management for wider tables and multi-disciplinary lists.

This configuration pairs a Luminox 300 compact head with a Luminox 500 head for variable Narrow and Broad focus on the primary position.

Key highlights

  • Dual dome configuration for overlapping coverage
  • Luminox 300 and Luminox 500 head pairing
  • 100,000 lux (300) and 160,000 lux (500) at 1 m
  • Variable Narrow and Broad focus on Luminox 500
  • 60,000+ hour LED service life

Specifications

Product specifications
ParameterDome 1: Luminox 300Dome 2: Luminox 500
Max. Central illumination at 1 m100,000 lux160,000 lux
Light field diameter at 1 m (d10)250 mm ± 10%200 mm - 300 mm ± 10%
Light head diameter300 mmup to 600 mm
Depth of illumination at 20% (L1+L2)1300 mm1200 mm
Depth of illumination at 60% (L1+L2)800 mm750 mm
Correlated Color Temperature (CCT)6500 K ± 500 K3500 K - 5500 K
Color Rendering Index (CRI)> 90> 92
Power consumption< 25 W< 50 W
Focus ModeFixedVariable (Narrow, Broad)
Dimming Range0 – 100 %0 – 100 %
No. of LEDs1230
LED Service Life60,000 hours60,000 hours
Supply Voltage230 V ±10%, 50 Hz ±5%230 V ±10%, 50 Hz ±5%
Power Supply Unit24 V DC ±10%, 4.2 A max24 V DC ±10%, 4.2 A max
Operating Temperature5°C - 40°C5°C - 40°C
Humidity10% - 90%10% - 90%
  • The specifications mentioned above can be used for wall, single, double, triple and mobile versions of Luminox series.
  • All values conform to IEC 60601-2-41 and are subject to a general industrial tolerance ± 5%.
  • In accordance with IEC 60601-2-41, the maximum illuminance must not exceed 160,000 lux.
  • Focus Mode: Variable = Narrow and Broad mode (Luminox 500). Luminox 300 uses fixed focus.
  • Custom development available for programme-specific mount and interface requirements.