Cresta Touch
Cresta I+sT
The Definitive Electrosurgical System - Engineered for Precision.
The CRESTA I+sT is an advanced RF energy generator platform engineered for precision electrosurgery. It delivers exceptional performance across monopolar, bipolar, and vessel sealing modes.
With state-of-the-art technologies including ContactIQ, SureBind, SafeZ, and TURiCUT, the CRESTA I+sT ensures maximum patient safety, consistent tissue effects, and superior surgical outcomes.
Key highlights
- Dual-Mode Operation
- Independent Activation
- Vessel seals up to 7 mm
- Consistent Tissue Effect
- Real time Tissue sensing Technology
- Underwater Monopolar cutting
- Return Electrode Monitoring
- RF Cut mode
- Intuitive 7" Touchscreen Control
- State-of-the-Art High-performance ARM32 Microcontroller (Sampling Rate >10,000/sec)
- Program Manager
- Self - Diagnosis
- Warnings
Specifications
| Parameter | Value |
|---|---|
| Outputs | Two Monopolar, One Bipolar, One Vessel Sealer |
| Dual-Mode Operation | Yes |
| Vessel Sealer | Yes |
| Saline Bipolar | NA |
| Pure Cut | 300 W |
| Vessel Sealer (High) | 150 W |
| Vessel seal diameter | up to 7 mm |
| Display | 7" Touchscreen |
| Weight | <10kg |
| Dimension (W x D x H) | 410mm x 160mm x 480mm |
| Input Supply voltage & frequency | 230V ± 10%, 50Hz |
| Operating Temperature range | +10°C to 40°C |
| Operating Relative Humidity | 15% to 85% non-condensing |
| Operating Atmospheric pressure | 700 to 1060 millibars |
| Transport & Storage Temp range | -10°C to +60°C |
| Transport & Storage Humidity | 15% to 90% |
| Transport & Storage Pressure | 500 to 1060 millibars |
Power modes
| Mode | Max Settings | Vpk-pk | Rated Load | Max Power in Watts | Crest Factor |
|---|---|---|---|---|---|
| MONOPOLAR CUT / Pure Cut | 300 | 3100 V | 500 Ω | 300 W | 1.4 |
| MONOPOLAR CUT / Blend 1 | 250 | 3300 V | 500 Ω | 250 W | 2.0 |
| MONOPOLAR CUT / Blend 2 | 200 | 3200 V | 500 Ω | 200 W | 2.3 |
| MONOPOLAR CUT / Blend 3 | 200 | 3600 V | 500 Ω | 200 W | 2.6 |
| MONOPOLAR COAG / Spray | 80 | 6000 V | 300 Ω | 80 W | 6.5 |
| MONOPOLAR COAG / Fulgurate | 120 | 6600 V | 300 Ω | 120 W | 5.0 |
| MONOPOLAR COAG / Desiccate 1 | 150 | 3100 V | 500 Ω | 150 W | 2.0 |
| MONOPOLAR COAG / Desiccate 2 | 150 | 2600 V | 500 Ω | 150 W | 1.8 |
| BIPOLAR / Micro | 80 | 650 V @ 500 Ω | 100 Ω | 80 W | 1.4 |
| BIPOLAR / Macro | 80 | 750 V @ 500 Ω | 100 Ω | 80 W | 1.6 |
| BIPOLAR / Cut | 120 | 800 V @ 500 Ω | 100 Ω | 120 W | 1.4 |
| VESSEL SEALER / Low | L2 | 300 V @ 150 Ω | - | 110 W | 1.4 |
| VESSEL SEALER / Average | A2 | 350 V @ 150 Ω | - | 130 W | 1.4 |
| VESSEL SEALER / High | H2 | 450 V @ 150 Ω | - | 150 W | 1.4 |
- The ratings mentioned above are maximum ratings.
- All output power values are available in ±20%; peak voltage and crest factors change accordingly.
- Power can be lowered during surgery by selecting appropriate settings.
- Always use the lowest power setting that achieves the desired surgical effect.
- Vessel Sealer power is modulated.