EPMD, Eclipse Prism Medical Devices
Cresta I+sT
Cresta Touch

Cresta I+sT

The Definitive Electrosurgical System - Engineered for Precision.

The CRESTA I+sT is an advanced RF energy generator platform engineered for precision electrosurgery. It delivers exceptional performance across monopolar, bipolar, and vessel sealing modes.

With state-of-the-art technologies including ContactIQ, SureBind, SafeZ, and TURiCUT, the CRESTA I+sT ensures maximum patient safety, consistent tissue effects, and superior surgical outcomes.

Key highlights

  • Dual-Mode Operation
  • Independent Activation
  • Vessel seals up to 7 mm
  • Consistent Tissue Effect
  • Real time Tissue sensing Technology
  • Underwater Monopolar cutting
  • Return Electrode Monitoring
  • RF Cut mode
  • Intuitive 7" Touchscreen Control
  • State-of-the-Art High-performance ARM32 Microcontroller (Sampling Rate >10,000/sec)
  • Program Manager
  • Self - Diagnosis
  • Warnings

Specifications

Product specifications
ParameterValue
OutputsTwo Monopolar, One Bipolar, One Vessel Sealer
Dual-Mode OperationYes
Vessel SealerYes
Saline BipolarNA
Pure Cut300 W
Vessel Sealer (High)150 W
Vessel seal diameterup to 7 mm
Display7" Touchscreen
Weight<10kg
Dimension (W x D x H)410mm x 160mm x 480mm
Input Supply voltage & frequency230V ± 10%, 50Hz
Operating Temperature range+10°C to 40°C
Operating Relative Humidity15% to 85% non-condensing
Operating Atmospheric pressure700 to 1060 millibars
Transport & Storage Temp range-10°C to +60°C
Transport & Storage Humidity15% to 90%
Transport & Storage Pressure500 to 1060 millibars

Power modes

Power modes
ModeMax SettingsVpk-pkRated LoadMax Power in WattsCrest Factor
MONOPOLAR CUT / Pure Cut3003100 V500 Ω300 W1.4
MONOPOLAR CUT / Blend 12503300 V500 Ω250 W2.0
MONOPOLAR CUT / Blend 22003200 V500 Ω200 W2.3
MONOPOLAR CUT / Blend 32003600 V500 Ω200 W2.6
MONOPOLAR COAG / Spray806000 V300 Ω80 W6.5
MONOPOLAR COAG / Fulgurate1206600 V300 Ω120 W5.0
MONOPOLAR COAG / Desiccate 11503100 V500 Ω150 W2.0
MONOPOLAR COAG / Desiccate 21502600 V500 Ω150 W1.8
BIPOLAR / Micro80650 V @ 500 Ω100 Ω80 W1.4
BIPOLAR / Macro80750 V @ 500 Ω100 Ω80 W1.6
BIPOLAR / Cut120800 V @ 500 Ω100 Ω120 W1.4
VESSEL SEALER / LowL2300 V @ 150 Ω-110 W1.4
VESSEL SEALER / AverageA2350 V @ 150 Ω-130 W1.4
VESSEL SEALER / HighH2450 V @ 150 Ω-150 W1.4
  • The ratings mentioned above are maximum ratings.
  • All output power values are available in ±20%; peak voltage and crest factors change accordingly.
  • Power can be lowered during surgery by selecting appropriate settings.
  • Always use the lowest power setting that achieves the desired surgical effect.
  • Vessel Sealer power is modulated.