Cresta Touch
Cresta BT
The Definitive Electrosurgical System - Engineered for Precision.
The CRESTA BT is an advanced RF energy generator platform engineered for precision electrosurgery. It delivers exceptional performance across monopolar and bipolar modes with consistent tissue effects.
With state-of-the-art technologies including ContactIQ, SafeZ, and TURiCUT, the CRESTA BT ensures real-time tissue sensing, consistent tissue effects, and superior surgical outcomes.
Key highlights
- Underwater Monopolar cutting
- Return Electrode Monitoring
- RF Cut mode
- Consistent Tissue Effect
- Real time Tissue sensing Technology
- Intuitive 7" Touchscreen Control
- State-of-the-Art High-performance ARM32 Microcontroller (Sampling Rate >10,000/sec)
- Program Manager
- Self - Diagnosis
- Warnings
Specifications
| Parameter | Value |
|---|---|
| Outputs | One Monopolar, One Bipolar |
| Vessel Sealer | NA |
| Saline Bipolar | NA |
| Pure Cut | 300 W |
| Blend 1 | 250 W |
| Spray | 80 W |
| Bipolar Cut | 120 W |
| Display | 7" Touchscreen |
| Underwater Monopolar | Yes (TURiCUT) |
| Return Electrode Monitoring | Yes (ContactIQ) |
| Weight | <10kg |
| Dimension (W x D x H) | 410mm x 160mm x 480mm |
| Input Supply voltage & frequency | 230V ± 10%, 50Hz |
| Operating Temperature range | +10°C to 40°C |
| Operating Relative Humidity | 15% to 85% non-condensing |
| Operating Atmospheric pressure | 700 to 1060 millibars |
| Transport & Storage Temp range | -10°C to +60°C |
| Transport & Storage Humidity | 15% to 90% |
| Transport & Storage Pressure | 500 to 1060 millibars |
Power modes
| Mode | Max Settings | Vpk-pk | Rated Load | Max Power in Watts | Crest Factor |
|---|---|---|---|---|---|
| MONOPOLAR CUT / Pure Cut | 300 | 3100 V | 500 Ω | 300 W | 1.4 |
| MONOPOLAR CUT / Blend 1 | 250 | 3300 V | 500 Ω | 250 W | 2.0 |
| MONOPOLAR CUT / Blend 2 | 200 | 3200 V | 500 Ω | 200 W | 2.3 |
| MONOPOLAR CUT / Blend 3 | 200 | 3600 V | 500 Ω | 200 W | 2.6 |
| MONOPOLAR COAG / Spray | 80 | 6000 V | 300 Ω | 80 W | 6.5 |
| MONOPOLAR COAG / Fulgurate | 120 | 6600 V | 300 Ω | 120 W | 5.0 |
| MONOPOLAR COAG / Desiccate 1 | 150 | 3100 V | 500 Ω | 150 W | 2.0 |
| MONOPOLAR COAG / Desiccate 2 | 150 | 2600 V | 500 Ω | 150 W | 1.8 |
| BIPOLAR / Micro | 80 | 650 V @ 500 Ω | 100 Ω | 80 W | 1.4 |
| BIPOLAR / Macro | 80 | 750 V @ 500 Ω | 100 Ω | 80 W | 1.6 |
| BIPOLAR / Cut | 120 | 800 V @ 500 Ω | 100 Ω | 120 W | 1.4 |
- The ratings mentioned in the above are maximum ratings.
- All the above output power are available in ±20%, accordingly the peak voltage and crest factors also change.
- The power can be lower to meet the requirements during surgery by selecting appropriate setting.
- Always use the lowest power setting that achieves the desired surgical effect.