Cresta Touch
Cresta B+T
The Definitive Electrosurgical System - Engineered for Precision.
The CRESTA B+T is an advanced RF energy generator platform engineered for precision electrosurgery. It delivers exceptional performance across monopolar and bipolar modes with consistent tissue effects.
With state-of-the-art technologies including ContactIQ, SafeZ, and TURiCUT, the CRESTA B+T ensures real-time tissue sensing, consistent tissue effects, and superior surgical outcomes.
Key highlights
- Dual-Mode Operation
- Independent Activation
- Underwater Monopolar cutting
- Return Electrode Monitoring
- RF Cut mode
- Consistent Tissue Effect
- Real time Tissue sensing Technology
- Intuitive 7" Touchscreen Control
- State-of-the-Art High-performance ARM32 Microcontroller (Sampling Rate >10,000/sec)
- Program Manager
- Self - Diagnosis
- Warnings
Specifications
| Parameter | Value |
|---|---|
| Outputs | Two Monopolar, One Bipolar |
| Dual-Mode Operation | Yes |
| Vessel Sealer | NA |
| Saline Bipolar | NA |
| Pure Cut | 300 W |
| Blend 1 | 250 W |
| Spray | 80 W |
| Bipolar Cut | 120 W |
| Display | 7" Touchscreen |
| Weight | <10kg |
| Dimension (W x D x H) | 410mm x 160mm x 480mm |
| Input Supply voltage & frequency | 230V ± 10%, 50Hz |
| Operating Temperature range | +10°C to 40°C |
| Operating Relative Humidity | 15% to 85% non-condensing |
| Operating Atmospheric pressure | 700 to 1060 millibars |
| Transport & Storage Temp range | -10°C to +60°C |
| Transport & Storage Humidity | 15% to 90% |
| Transport & Storage Pressure | 500 to 1060 millibars |
Power modes
| Mode | Max Settings | Vpk-pk | Rated Load | Max Power in Watts | Crest Factor |
|---|---|---|---|---|---|
| MONOPOLAR CUT / Pure Cut | 300 | 3100 V | 500 Ω | 300 W | 1.4 |
| MONOPOLAR CUT / Blend 1 | 250 | 3300 V | 500 Ω | 250 W | 2.0 |
| MONOPOLAR CUT / Blend 2 | 200 | 3200 V | 500 Ω | 200 W | 2.3 |
| MONOPOLAR CUT / Blend 3 | 200 | 3600 V | 500 Ω | 200 W | 2.6 |
| MONOPOLAR COAG / Spray | 80 | 6000 V | 300 Ω | 80 W | 6.5 |
| MONOPOLAR COAG / Fulgurate | 120 | 6600 V | 300 Ω | 120 W | 5.0 |
| MONOPOLAR COAG / Desiccate 1 | 150 | 3100 V | 500 Ω | 150 W | 2.0 |
| MONOPOLAR COAG / Desiccate 2 | 150 | 2600 V | 500 Ω | 150 W | 1.8 |
| BIPOLAR / Micro | 80 | 650 V @ 500 Ω | 100 Ω | 80 W | 1.4 |
| BIPOLAR / Macro | 80 | 750 V @ 500 Ω | 100 Ω | 80 W | 1.6 |
| BIPOLAR / Cut | 120 | 800 V @ 500 Ω | 100 Ω | 120 W | 1.4 |
- The ratings mentioned in the above are maximum ratings.
- All the above output power are available in ±20%, accordingly the peak voltage and crest factors also change.
- The power can be lower to meet the requirements during surgery by selecting appropriate setting.
- Always use the lowest power setting that achieves the desired surgical effect.