EPMD, Eclipse Prism Medical Devices
Cresta Ws
Cresta Digital

Cresta Ws

Full modality set on split ports including ReSaliX saline bipolar.

Cresta Ws is the Digital full-platform configuration: monopolar, bipolar, SureBind vessel sealing, and ReSaliX saline bipolar on split bipolar and vessel sealer ports.

The brochure power table confirms vessel sealer and saline bipolar wattages on this variant. SafeTis, ContactIQ, and TURiCUT-capable monopolar modes cover urology, gynaecology, and general surgery under saline or open conditions.

Key highlights

  • Monopolar, bipolar, vessel sealer, saline bipolar
  • Split port configuration
  • ReSaliX saline activation
  • 7-segment LED interface
  • Saline bipolar Cut 300 W
  • Ordering code LF90127

Specifications

Product specifications
ParameterValue
OutputsOne Monopolar, One Bipolar, Vessel Sealer, Saline Bipolar
Display7-segment LED
Port configurationSplit (Bipolar and Vessel Sealer)
Vessel SealerYes
Saline BipolarYes
Saline Bipolar Cut300 W
Ordering codeLF90127
Weight<10kg
Dimension (W x D x H)410mm x 160mm x 480mm
Input Supply voltage & frequency230V ± 10%, 50Hz
Operating Temperature range+10°C to 40°C
Operating Relative Humidity15% to 85% non-condensing
Operating Atmospheric pressure700 to 1060 millibars
Transport & Storage Temp range-10°C to +60°C
Transport & Storage Humidity15% to 90%
Transport & Storage Pressure500 to 1060 millibars

Power modes

Power modes
ModeMax SettingsVpk-pkRated LoadMax Power in WattsCrest Factor
MONOPOLAR CUT / Pure Cut3003100 V500 Ω300 W1.4
MONOPOLAR CUT / Blend 12503300 V500 Ω250 W2.0
MONOPOLAR COAG / Spray806000 V300 Ω80 W6.5
BIPOLAR / Cut120800 V @ 500 Ω100 Ω120 W1.4
VESSEL SEALER / LowL2300 V @ 150 Ω-110 W1.4
VESSEL SEALER / HighH2450 V @ 150 Ω-150 W1.4
SALINE BIPOLAR / CutL61500 V100 Ω300 W1.4
SALINE BIPOLAR / CoagL6950 V100 Ω250 W1.4
  • The ratings mentioned above are maximum ratings.
  • All output power values are available in ±20%; peak voltage and crest factors change accordingly.
  • Power can be lowered during surgery by selecting appropriate settings.
  • Always use the lowest power setting that achieves the desired surgical effect.
  • Vessel Sealer power is modulated.