EPMD, Eclipse Prism Medical Devices
Cresta Wc
Cresta Digital

Cresta Wc

Full modality set on a common bipolar and vessel sealer port.

Cresta Wc mirrors Cresta Ws clinically but uses a common port for bipolar and vessel sealer connections, with monopolar and ReSaliX saline bipolar outputs unchanged.

Power-table ratings include vessel sealer and saline bipolar modes. The 7-segment LED console keeps mode and wattage visible under low-light OR conditions.

Key highlights

  • Monopolar, bipolar, vessel sealer, saline bipolar
  • Common port configuration
  • ReSaliX saline activation
  • 7-segment LED interface
  • Saline bipolar Cut 300 W
  • Ordering code LF90128

Specifications

Product specifications
ParameterValue
OutputsOne Monopolar, One Bipolar, Vessel Sealer, Saline Bipolar
Display7-segment LED
Port configurationCommon (Bipolar and Vessel Sealer)
Vessel SealerYes
Saline BipolarYes
Saline Bipolar Cut300 W
Ordering codeLF90128
Weight<10kg
Dimension (W x D x H)410mm x 160mm x 480mm
Input Supply voltage & frequency230V ± 10%, 50Hz
Operating Temperature range+10°C to 40°C
Operating Relative Humidity15% to 85% non-condensing
Operating Atmospheric pressure700 to 1060 millibars
Transport & Storage Temp range-10°C to +60°C
Transport & Storage Humidity15% to 90%
Transport & Storage Pressure500 to 1060 millibars

Power modes

Power modes
ModeMax SettingsVpk-pkRated LoadMax Power in WattsCrest Factor
MONOPOLAR CUT / Pure Cut3003100 V500 Ω300 W1.4
MONOPOLAR CUT / Blend 12503300 V500 Ω250 W2.0
MONOPOLAR COAG / Spray806000 V300 Ω80 W6.5
BIPOLAR / Cut120800 V @ 500 Ω100 Ω120 W1.4
VESSEL SEALER / LowL2300 V @ 150 Ω-110 W1.4
VESSEL SEALER / HighH2450 V @ 150 Ω-150 W1.4
SALINE BIPOLAR / CutL61500 V100 Ω300 W1.4
SALINE BIPOLAR / CoagL6950 V100 Ω250 W1.4
  • The ratings mentioned above are maximum ratings.
  • All output power values are available in ±20%; peak voltage and crest factors change accordingly.
  • Power can be lowered during surgery by selecting appropriate settings.
  • Always use the lowest power setting that achieves the desired surgical effect.
  • Vessel Sealer power is modulated.